Title | SKU | Price |
---|---|---|
Default Title | BE:BRC15110U |
$ 3,507
|
Designed so that test results are in line with international standards DIN 12880.
P.I.D temperature controller provides accurate and reliable temperature control.
Automatic calibration function, be able to adapt to a variety of laboratory environment.
A spread product line meets a variety of needs
Wide range of options to meet a variety of experimental needs.
Temperature/Capacity/Liquid Flow/Pressure/Control Mode/Internal Circulation/External Circulation.
Convenient design for easy maintenance
Round curved inner angle is easy to clean.
Inner chamber is made from corrosion-resistant mirror stainless steel.
Preset On/Off function.
Built-in power interruption protection function, automatic run after power interruption.
Independent Over-temperature protection meets DIN 12880 International standard requirements.
Temperature deviation alarm.
Storage Tank Capacity(L) | 10 | ||||||||
---|---|---|---|---|---|---|---|---|---|
Lowest Temp Without Load(℃) | ﹣20℃ | ||||||||
Temperature Range(℃) | ﹣20℃~20℃ | ||||||||
Best Operating Temp(℃) | ≤35℃ | ||||||||
Relative Humidity(%RH) | ≤85%RH | ||||||||
Temperature stability | ±2℃ | ||||||||
Cooling Capacity (w) |
|
||||||||
Refrigerant | R404A | ||||||||
Temperature Sensor | pt100 | ||||||||
Screen | LCD Screen | ||||||||
Safety Protection | Compressor delay start、Leakage circuit breaker,Over current/ Overvoltage Protection | ||||||||
Power(W)(whole chiller) | 650 | ||||||||
Power Requirement | AC 115V/60Hz | ||||||||
Recirculating Pump Power(W) | 25 | ||||||||
Pump Flow Rate(L/min) | 27 | ||||||||
Max. Pump Head(m) | 3 | ||||||||
Inlet/Outlet Pipe Diameter(mm) | φ18 | ||||||||
Inlet/Outlet Connector Diameter(mm) | φ12 | ||||||||
Noise | ≤55 | ||||||||
Weight(Kg) | 60 | ||||||||
Liquid tank Dimension(mm) | φ250mm*250mm | ||||||||
Exterior Dimension(mm)(including casters) | 400*540*790 | ||||||||
Adaptated Rotavaps(Capacity*Units) | 5L*2 Or 10L*1 | ||||||||
Sku Numbers | BRC15110U |
The specifications show the lowest temperature without load for the BRC-10A Recirculating Chiller is -20°C, but also says it has 1000 W cooling capacity at that temperature? Can it hold -20°C with a load on it?
Being states that under their testing the BRC-10A chiller is 60% efficient at -20°C, so it would be able to maintain -20°C with a load of no more than 600 W. -20°C will be the lowest temperature with or without load, as it is the lowest setpoint temperature for the chiller.
How is the power specification for the whole chiller lower than the maximum cooling capacity?
This specification is given as a normal operating power consumption. The actual power consumption will vary based on usage and may be significantly higher in demanding applications.
3692 BEING 2021-03-28